Introduction
The Pulmonary Artery Catheter (PAC) Swan-Ganz, despite the widespread rumours regarding
the possible complications that correlate with its insertion, such as the pulmonary
artery rupture, is still - in the experienced hands - an essential tool of hemodynamic
monitoring in cardiac surgery. In combination with Transoesophageal Echocardiography
(TOE) though, they offer to the Anaesthesiologist an integrated profile of pressure,
volume, flow velocity and anatomical information. However, there can still be occasions
where the above mentioned hemodynamic and non-hemodynamic data are not enough to guide
the surgical decision-making process intraoperatively. In this case report, we present
the contribution of the difficulty in advancing the PAC into the Right Atrium (RA),
as an indication of Superior Vena Cava (SVC) stenosis, and the need to return to CardioPulmonary
Bypass (CPB) to repair it, during a surgery of totally endoscopic Atrial Septal Defect
(ASD) closure.

Methods
A 40yr old female patient is undergoing totally endoscopic cardiac surgery with CPB
for the closure of a Superior Sinus Venosus ASD with simultaneous anomalous drainage
of the Right Upper Pulmonary Vein (RUPV) into the SVC. The anaesthetic monitoring
includes arterial line, PAC and TOE, which confirms the preoperative findings (Figure
1,2,3), while for the needs of the operation the Anaesthesiologist inserts also a
CPB cannula in the Right Internal Jugular Vein. Later on, during the phase that the
surgeon is advancing the second venous CPB cannula through the femoral vein and the
Inferior Vena Cava (IVC) into the RA, the PAC is being withdrawn from its wedge position
and is left high inside the SVC, showing the Central Venous Pressure (CVP).
Results
After the closure of the ASD and the flow diversion of the RUPV into the Left Atrium
with a bovine pericardial patch, deairing is taking place, the restoration of the
ASD is being confirmed with the TOE and successful CPB weaning occurs. A subsequent
TOE scan shows turbulence inside the SVC with the colour Doppler (Figure 4), indicative
of stenosis, while the CVP is 15mmHg, and the dilemma that is coming up is if the
stenosis is clinically significant or not. However, at the same time it is impossible
to readvance the PAC into the RA, an indication of severe SVC stenosis, and the team
decides to return to CPB to repair it. After the placement of a second pericardial
patch that widens the RA-SVC junction on a beating heart, laminar flow is being depicted
inside the SVC (Figure 5), the PAC is being seamlessly floated into the RA (Figure
6) until its wedge position, and the CVP is 8mmHg. The short and the long-term postoperative
periods were uncomplicated.
Discussion
The intraoperative use of the PAC in the current case report, and the difficulty in
readvancing it more specifically, proved to be crucial in making an undoubtedly critical
decision towards repairing a stenosis that could have been overlooked, if we had simply
relied on the CVP, and become even fatal in case of delayed diagnosis. The mechanical
contribution of the PAC has been underlined.
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Journal of Cardiothoracic and Vascular AnesthesiaAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
Article info
Identification
Copyright
© 2021 Published by Elsevier Inc.