Objective
To compare the reliability of cardiac index (CI) and stroke-volume variation (SVV)
measured by the pulse-wave transit-time (PWTT) method using estimated continuous cardiac
output (esCCO) technique with conventional pulse-contour analysis after off-pump coronary
artery bypass grafting (OPCAB).
Design
A single-center, prospective, observational study.
Setting
At a 1,000-bed university hospital.
Participants
A total of 21 patients were enrolled after elective OPCAB.
Interventions
The study authors performed a method comparison study with simultaneous measurement
of CI and SVV based on the esCCO technique (CIesCCO and esSVV, correspondingly) and pulse-contour analysis (CIPCA and SVVPCA, correspondingly). As a secondary analysis, they also assessed the trending ability
of CIesCCO versus CIPCA.
Measurements and Main Results
The authors analyzed 178 measurement pairs for CI, and 174 pairs for SVV during the
10 study stages. The mean bias between CIesCCO and CIPCA was 0.06 L min/m2, with limits of agreement of ± 0.92 L min/m2 and a percentage error (PE) of 35.3%. The analysis of the trending ability of CI
measured by PWTT revealed a concordance rate of 70%. The mean bias between esSVV and
SVVPCA was –6.1%, with limits of agreement of ± 15.5% and a PE of 137%.
Conclusions
The overall performance of CIesCCO and esSVV versus CIPCA and SVVPCA is not clinically acceptable. A further improvement of the PWTT algorithm may be
required for an accurate and precise assessment of CI and SVV.
Key Words
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to Journal of Cardiothoracic and Vascular AnesthesiaAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- New developments in haemodynamic monitoring.J Cardiothorac Vasc Anesth. 2019; 33: 67-72
- Best practice & research clinical anaesthesiology: Advances in haemodynamic monitoring for the perioperative patient: Perioperative cardiac output monitoring.Best Pract Res Clin Anaesthesiol. 2019; 33: 139-153
- Perioperative blood pressure management.Anesthesiology. 2021; 134: 250-261
- Pulse wave analysis.in: Kirov M Kuzkov V Saugel B Advanced Haemodynamic Monitoring: Basics and New Horizons. Springer, Cham, Switzerland2021: 69-78
- Can we obtain a noninvasive and continuous estimation of cardiac output? Comparison between three noninvasive methods.Int Heart J. 2013; 54: 395-400
- Non-invasive estimation of stroke volume index and its variation by using pulse wave transit time.Eur J Anaesthesiol. 2010; 27: 65
- Pulse-wave transit time with ventilator-induced variation for the prediction of fluid responsiveness.Acute Med Surg. 2020; 7: 10
- Accuracy of noninvasive estimated continuous cardiac output (esCCO) Compared to Thermodilution Cardiac Output: A pilot study in cardiac patients.J Cardiothorac Vasc Anesth. 2013; 27: 1128-1132
- Estimated continuous cardiac output based on pulse wave transit time in off-pump coronary artery bypass grafting: A comparison with transpulmonary thermodilution.J Clin Monit Comput. 2017; 31: 361-370
- A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors.J Clin Monit Comput. 2004; 18: 13-20
- Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: A comparison with intermittent bolus thermodilution cardiac output.Anesth Analg. 2012; 115: 82-87
- Cardiac output and stroke volume variation measured by the pulse wave transit time method: A comparison with an arterial pressure-based cardiac output system.J Clin Monit Comput. 2019; 33: 385-392
- Prediction of fluid responsiveness by means of stroke volume variation measured by pulse wave transit time-based cardiac output monitoring.Toho J Med. 2020; 6: 41-47
- Validation of cardiac output monitoring based on uncalibrated pulse contour analysis vs transpulmonary thermodilution during off-pump coronary artery bypass grafting.Br J Anaesth. 2014; 112: 1024-1031
- Perioperative cardiovascular monitoring of high-risk patients: A consensus of 12.Crit Care. 2015; 19: 224
- A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output.Anesth Analg. 2010; 111: 1180-1192
- Comparison of esCCO and transthoracic echocardiography for non-invasive measurement of cardiac output intensive care.Br J Anaesth. 2012; 109: 879-886
- The ability of esCCO and ECOM monitors to measure trends in cardiac output during alveolar recruitment maneuver after cardiac surgery: A comparison with the pulmonary thermodilution method.Anesth Analg. 2015; 121: 383-391
- Can changes in arterial pressure be used to detect changes in cardiac output during volume expansion in the perioperative period?.Anesthesiology. 2012; 117: 1165-1174
- Accuracy of stroke volume variation in predicting fluid responsiveness: A systematic review and meta-analysis.J Anesth. 2011; 25: 904-916
- Stroke volume variation for assessment of cardiac responsiveness to volume loading in mechanically ventilated patients after cardiac surgery.Intensive Care Med. 2002; 28: 392-398
- Prediction of fluid responsiveness in mechanically ventilated cardiac surgical patients: The performance of seven different functional haemodynamic parameters.BMC Anesthesiol. 2018; 18: 55
- Dynamic indices.in: Kirov M Kuzkov V Saugel B Advanced Haemodynamic Monitoring: Basics and New Horizons. Springer, Cham, Switzerland2021: 149-159
- Prediction of fluid responsiveness: An update.Ann Intensive Care. 2016; 6: 111
- The PiCCO monitor.Int Anesthesiol Clin. 2010; 48: 57-85
- Pressure waveform analysis.Anesth Analg. 2018; 126: 1930-1933
Article info
Publication history
Published online: February 11, 2023
Footnotes
This study was partially supported by a grant from Nihon Kohden (Japan).
Identification
Copyright
© 2023 Elsevier Inc. All rights reserved.